64
© Servomex Group Limited. 2020
07931001B revision 6
To set the relay coil state
Figure 4-25: Relay coil state menu
Note:
You can specify the relay coil state (energized or de-energized) to meet any
particular application requirement. For example, Relay 1 coil is de-energized
during an active status when a measurement or analyzer fault is detected.
Therefore, if the power to the analyzer fails, or the relay cable is disconnected, a
fault is raised because this is electrically the same as if Relay 1 is de-energized.
This is a ‘fail
-
safe’ situation. In some applications it is preferable that the relay is
energized in an active status.
Menu
Set Up
Relay Setup
Relay Setup
Calibrate
Manage Analyser
Relay
Relay 1 Event 1
Data Log
mA Input
Alarms
mA Output
1
System Fault
Status
Relay Setup
Set Up
Filtering
Measurements
Unit Select
Service
X-Interference
Clipping
User Settings
Relay Setup
Relay Setup
Relay 1 active
Relay 1 active
Energised
Deenergised
Deenergised
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...