86
© Servomex Group Limited. 2020
07931001B revision 6
6.4 Reconfigure settings for off-line calibration
Adjust the ‘physical set up’ to suit the calibration cell environment. Path length,
pressure source, temp source, analogue I/P.
Note:
If the analyzer has been set with a long measurement filter the filter
should be reset to an appropriate time for example, 10 seconds to avoid a
poor calibration result.
Alternatively, if an offline calibration settings file has already been saved. Load the
saved file as described in 0
Figure 6-2: Physical setup sequential menu
Check photodiode DC voltage and signal amplitude. Adjust if required. Ref. 4.11
Calibration cell path length to be entered: 250mm, 500mm, 750mm or 1000mm
Menu
Manage Analyser
Calibrate
Level 3
Level 4
Level 5
Level 6
Level 7
Level 8
Level 9
Data log
Alarms
Status
Set up
Measurement
Service
Enabled
Disabled
Manage Current Config
Service Mode
Phase/Gain Settings
Measurement Signal
Physical Setup
Measurement
Selection
Purge Compensation
Proc:Path Length
0.000
Transmitter Enclosure
User defined 0-4
mA Input 1 or 2
User defined 0-4
select option
select option
Proc:Pressure Source
Proc:Temperature Source
mA Input 1 or 2
select option
Receiver Enclosure
Proc:Pressure Source
Offset or value
Edit
Transmitter Enclosure
edit value
edit value
Delete Config
edit value
Detailed set up
For Servomex trained Service operators only
Save Current Config
Restore Config
Proc:Temp. Offset
Offset or value
Edit
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...