07931001B revision 6
© Servomex Group Limited. 2020
7
2. Introduction
2.1 Product description
The Laser 3 Plus is an advanced gas analyzer. It comprises a transmitter unit (TU)
and a receiver unit (RU), with an optional AC to DC power supply, ancillary
equipment and sample cell.
The laser radiation detected by a photodetector is amplified and transferred to the
transmitter unit through a multi-core cable which connects the receiver and
transmitter electronics. The same cable transfers the required power from the TU to
the RU. Connectors at each end of the cable, and a removable gland plate, allow
removal of the TU and RU for calibration or maintenance.
Flange mounts and window purge assemblies are provided to mount the transmitter
and receiver to the process flanges. Quick-connect connections between the
instrument and mounting assembly allows easy removal of the transmitter and
receiver for calibration or maintenance.
For continuous in-situ applications, the Laser 3 Plus is installed directly across
stacks, ducts and reactors typically employing path lengths of 0.1 to 25m (3.94" -
82ft.).
Figure 2-1: Laser 3 Plus example installation (exploded view)
1
Receiver Unit
6
Process gas
2
O-ring
7
Adjustable mount (2 variants)
3
Fixed mount or adjustable
8
Transmitter Unit
4
Gasket or O-ring
9
Non-visible laser beam path
5
Process pipe, stack, duct, etc.
10
Process flange and nozzle
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...