136
© Servomex Group Limited. 2020
07931001B revision 6
Refer to
Step 1
Switch on to light source in order to check operation before fitting then fit the
alignment tool light source assembly to the Analyzer flange (fixed or
adjustable) on the RU side of the process using 3 off M6 x 16 screws. Switch
the light source on
Step2
Fit the alignment tool eye piece assembly to the Analyzer flange (adjustable)
on the TU side of the process using 3 off M6x16 screws. Peel back the
protective boot on the ball joint assembly to reveal the adjusting screws (Figure
7-10)
Note:
Ensure all the M6 x 16 screws are tight before beginning the alignment
process.
Step 3
Loosen the adjustment nuts and screws by turning anti-clockwise.
Ref (Figure
Step 4
For correct alignment the light source must be in the center of the eye piece. If
the eye piece is off target adjust the TU/RU alignment assembly tighten the
adjustment screws to lock the assembly in position. This is easiest to do with
two hex keys, one to loosen and the other to tighten.
Note:
All adjusting screws have a right hand M6 thread and give an angular
resolution of 1.8º per screw turn. The ball joint may be adjusted by ±4º in each
direction
Step 5
When all adjustments are complete, ensure all screws are locked in place
Hint:
Taking care not to change the position of the assembly during steps 5 and 6
Step 6
Remove the alignment tool scope assembly and replace the boot immediately
Step 7
Switch off the alignment tool light source assembly and remove it from the RU
Step 8
If the analyzer has been supplied with an adjustable, rather than fixed
mounting assembly at the receiver end, fit the alignment tool light source
assembly to the analyzer flange on the TU side of the process as described in
step 1. Fit the alignment tool scope assembly to the RU side of the process as
described in step 2 and repeat steps 4 to 7, to align and secure the second
analyzer mounting assembly.
Note:
The adjustable receiver mount is adjusted by a combination of set screws and
mounting screws, see
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...