102
© Servomex Group Limited. 2020
07931001B revision 6
Step
No
Display
Description
All Good Path
Failure Path
28
Apply span-gas to the validation cell
and then press
to continue. Note
that you can press
immediately
after applying span-gas as the
analyzer itself will apply a flush time.
31
The operator is now informed
whether the validation has passed or
failed. A pass means that the
calculated span concentration was
within the permitted range.
32
The operator is now instructed to
purge the validation cell with zero-
gas.
Note that in case of the Combustion
O2 analyzer variant, the normal
condition for the purge might be
Instrument Air if Nitrogen is not
available at site. If that is the case
Nitrogen purge of the in-line gas cell
will be required, only for the
validation test, in order to produce a
change in Oxygen concentration
from the gas cell.
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...