148
© Servomex Group Limited. 2020
07931001B revision 6
Ensure connector
latches “click” and are engaged fully.
Before you remove the receiver unit cover, turn off the power to the
transmitter unit to ensure that there is no risk of exposure to the laser
beam.
The connection cable must have the following specification:
Cable type
Screened
Number of cores
8
Configuration of cores
Twisted pairs
Rating (current / voltage)
300 V
RMS
Characteristic impedance
100
Ω
Rating (temperature)
-20 to +75 °C (-4 to 167 °F)
Cable conductors
0.2 to 1.5 mm
2
(28 to 14 AWG)
Flammability
VW-1
Approvals
Relevant to local requirements
Cable external diameter
Within the range specified for the selected cable
gland
Maximum length
100 m
Note:
Under certain conditions of high ambient temperature a higher temperature rated
cable may be required.