Calibrate the mA outputs
Figure 4-14: mA output calibration and override menu
Supervisor level is required to calibrate the mA outputs. Select the mA output to be
calibrated, 1 or 2 and scroll down to calibrate screen. As soon as the mA output
calibrate screen is shown, the nominal mA output value is set to 20 mA:
Use your control/monitoring equipment (connected to the analyzer) to report the
actual output value.
Use the soft keys to increase or decrease the actual output value until your control /
monitoring equipment indicates 20 mA output.
When the mA output has been correctly calibrated, press the soft key, the mA output
service screen will then be displayed again.
Menu
Service
mA Output
mA Output 1 : 1 XXX
mA Output 1 : 1 XXX
Calibrate
mA Output
Override
mA Output
mA Output
Data Log
mA Input
Calibrate
2
Alarms
Relay Override
1
1
Status
Set Up
Measurements
Service
adjust
mA Output 1 : 1 XXX
mA Output 1 : 1 XXX
Calibrate
Calibrate
mA
20.000 mA
adjust
select
select
Menu
Service
mA Output
mA Output 1 : 1 XXX
mA Output 1 : 1 XXX
Calibrate
mA Output
Overide
mA Output
mA Output
Data Log
mA Input
Calibrate
2
Alarms
Relay Override
1
1
select
Status
Set Up
Measurements
Service
mA Output 1 : 1 XXX
mA Output 1 : 1 XXX
Override Value
Override Value
2.000 mA
2.000 mA
edit value
mA Output 1 : 1 XXX
mA Output 1 : 1 XXX
Override
Override
Yes
No
No
select
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...