07931001B revision 6
© Servomex Group Limited. 2020
97
Step
No
Display
Description
All Good Path
Failure Path
4
If the system has a fault currently
active, then a message will be
displayed saying that validation
cannot continue. It will then exit back
to the top level calibrate menu
If no fault is present, then enter the
operator password to continue
validation. This step will be skipped
if you are already logged in at
operator level or above. At this point
the analyzer will raise the Service In
Progress flag.
If the system is in warming mode,
then a message will be displayed
saying that validation cannot
continue.
5
Enter the pressure in the validation
cell. In this example the cell is at
atmospheric pressure.
6
Enter the temperature in the
validation cell
7
At this point you have the choice
whether to review and edit all the
validation settings or jump to the
start of the validation process.
Bypassing the settings will result in
the stored settings being used.
Press
to review or
to jump. If
you
select jump, the process will
immediately go to step 17.
8
Choose the span target value. This
will usually be the concentration of
the span-gas.
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...