To adjust DC Gain 1 and Gain 2 settings
Figure 4-32: Adjust gain and phase menu
Note:
The target photodiode DC level will depend on the application and site
characteristics. For example, dust levels and amount of beam
divergence for long path length applications.
The maximum DC gain level will depend on the gas measurement and
pathlength. These limits are set as part of the measurement
parameters.
Too high photodiode DC level will cause an XXXX error status and
incorrect readings
Level 1
Level 2
Level 3
Level 4
Level 5
Menu
Set Up
Manage Analyser
Manage Current Config
Phase/Gain Settings
Calibrate
Manage Analyser
Manage Current Config
Service Mode
Alignment LEDs
Data Log
mA Input
Save Current Config
Phase/Gain Settings
Alignment Lock Level
Alarms
mA Output
Restore Config
Measurement Signal
Photodiode DC level
Status
Relay Setup
Delete Config
Physical Setup
DC Gain 1
Set Up
Filtering
Detailed Setup
DC Gain 2
Measurements
Unit Select
0
Transmission Level
Service
X-Interference
Raw Signal
Clipping
AC Amplifier Gain
User Settings
ABurst
Adjust Burst Phase Cal
Ref TransCal
Phase/Gain Settings
Photodiode DC level
2.00V
Check
Phase/Gain Settings
DC Gain 1
120
Set (min 0 - max 255)
Phase/Gain Settings
DC Gain 2
40
Set (min 0 - max 255)
Содержание 07931001B
Страница 1: ...Part Number 07931001B Revision 6 Language ENGLISH UK US SERVOTOUGH Laser 3 Plus OPERATOR MANUAL...
Страница 2: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 4: ...Servomex Group Limited 2020 This page intentionally blank...
Страница 12: ...viii Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 14: ...x Servomex Group Limited 2020 07931001B revision 6 This page intentionally blank...
Страница 172: ...158 Servomex Group Limited 2020 07931001B revision 6 Figure 7 25 Installation ball joint adjustment screws...