MSRA configuration
R&S
®
FSWP-B1
46
User Manual 1177.5862.02 ─ 09
Drop-Out Time ← Trigger Source
Defines the time that the input signal must stay below the trigger level before triggering
again.
Remote command:
Trigger Offset ← Trigger Source
Defines the time offset between the trigger event and the start of the sweep for the
MSRA primary.
This setting is not available in MSRA secondary application measurement channels.
For secondary applications, use the
offset > 0:
Start of the sweep is delayed
offset < 0:
Sweep starts earlier (pre-trigger)
Maximum allowed range limited by the sweep time:
pretrigger
max
= sweep time
Remote command:
TRIGger[:SEQuence]:HOLDoff[:TIME]
Hysteresis ← Trigger Source
Defines the distance in dB to the trigger level that the trigger source must exceed
before a trigger event occurs. Setting a hysteresis avoids unwanted trigger events
caused by noise oscillation around the trigger level.
This setting is only available for "IF Power" trigger sources. The range of the value is
between 3
dB and 50
dB with a step width of 1
dB.
Remote command:
TRIGger[:SEQuence]:IFPower:HYSTeresis
Trigger Holdoff ← Trigger Source
Defines the minimum time (in seconds) that must pass between two trigger events.
Trigger events that occur during the holdoff time are ignored.
Remote command:
TRIGger[:SEQuence]:IFPower:HOLDoff
Slope ← Trigger Source
For all trigger sources except time, you can define whether triggering occurs when the
signal rises to the trigger level or falls down to it.
Remote command:
Trigger settings