Welcome to the MSRA operating mode
R&S
®
FSWP-B1
15
User Manual 1177.5862.02 ─ 09
1 = Channel information bar for the MSRA Primary
2 = Secondary application data coverage for each active secondary application
3 = Result display for MSRA Primary (for entire capture buffer)
4 = Channel information bar for secondary application with button to switch to secondary application tab
5 = Result display for secondary application (for analysis interval)
2.2.2
MSRA primary
The MSRA primary is the only channel that captures data. It also controls global con-
figuration settings for all secondary applications. The MSRA primary channel itself is
implemented as an I/Q Analyzer secondary application. The MSRA primary measure-
ment channel cannot be deleted or replaced.
The following figure shows the screen elements specific to the MSRA primary.
1
2
3
4
1 = Channel information bar for the MSRA Primary
2 = Data coverage for each active secondary application
3 = Result display for MSRA Primary (for entire capture buffer)
4 = Analysis line
Channel bar information
The channel bar shows the firmware and measurement information for data acquisition
and global configuration.
Table 2-1: Information displayed in the channel bar for the MSRA Primary
Ref Level
Reference level
Att
RF attenuation
Ref Offset
Reference level offset
Freq
Center frequency
Understanding the display information