Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
74
User Manual 1177.5862.02 ─ 09
INSTrument[:SELect]
<ChannelType> | <ChannelName>
This command activates a new channel with the defined channel type, or selects an
existing channel with the specified name.
Also see
●
Parameters:
<ChannelType>
Channel type of the new channel.
For a list of available channel types see
<ChannelName>
String containing the name of the channel.
Example:
INST IQ
Activates a channel for the I/Q Analyzer application (evaluation
mode).
INST 'MyIQSpectrum'
Selects the channel named 'MyIQSpectrum' (for example before
executing further commands for that channel).
Manual operation:
See
"AM/FM/PM Modulation Analysis"
See
See
See
"Vector Signal Analysis (VSA)"
See
"Selecting an MSRA secondary application"
See
SYSTem:PRESet:COMPatible
<OpMode>
This command defines the operating mode that is activated when you switch on the
R&S
FSWP or press the [PRESET] key.
For details on operating modes see
Chapter 4, "Applications and operating modes"
Parameters:
<OpMode>
SANalyzer
Defines Signal and Spectrum Analyzer operating mode as the
presetting.
OFF
Selects the phase noise application as the default application
(default value).
SYSTem:PRESet:CHANnel[:EXEC]
This command restores the default instrument settings in the current channel.
Use
INST:SEL
to select the channel.
Activating MSRA measurements