Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
90
User Manual 1177.5862.02 ─ 09
Parameters:
<TriggerLevel>
Range:
0.5 V to 3.5 V
*RST:
1.4 V
Default unit: V
Example:
TRIG:LEV 2V
Manual operation:
See
TRIGger[:SEQuence]:LEVel:IFPower
<TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths,
see the data sheet.
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IFP -30DBM
TRIGger[:SEQuence]:LEVel:IQPower
<TriggerLevel>
This command defines the magnitude the I/Q data must exceed to cause a trigger
event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel>
Range:
-130 dBm to 30 dBm
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IQP -30DBM
TRIGger[:SEQuence]:LEVel:RFPower
<TriggerLevel>
This command defines the power level the RF input must exceed to cause a trigger
event. Note that any RF attenuation or preamplification is considered when the trigger
level is analyzed. If defined, a reference level offset is also considered.
The input signal must be between 500
MHz and 8
GHz.
Parameters:
<TriggerLevel>
For details on available trigger levels and trigger bandwidths,
see the data sheet.
*RST:
-20 dBm
Default unit: DBM
Configuring MSRA measurements