Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
93
User Manual 1177.5862.02 ─ 09
<tp>
Selects the trigger port to which the output is sent.
2 = trigger port 2 (rear)
Parameters:
<Level>
HIGH
5 V
LOW
0 V
*RST:
LOW
Example:
OUTP:TRIG2:LEV HIGH
Manual operation:
See
OUTPut<up>:TRIGger<tp>:OTYPe
<OutputType>
This command selects the type of signal generated at the trigger output.
Suffix:
<up>
.
1..n
<tp>
Selects the trigger port to which the output is sent.
2 = trigger port 2 (rear panel)
Parameters:
<OutputType>
DEVice
Sends a trigger signal when the R&S
FSWP has triggered inter-
nally.
TARMed
Sends a trigger signal when the trigger is armed and ready for
an external trigger event.
UDEFined
Sends a user-defined trigger signal. For more information, see
*RST:
DEVice
Manual operation:
See
OUTPut<up>:TRIGger<tp>:PULSe:IMMediate
This command generates a pulse at the trigger output.
Suffix:
<up>
.
Selects the trigger port to which the output is sent.
2 = trigger port 2 (rear)
<tp>
1..n
Manual operation:
See
Configuring MSRA measurements