MSRA configuration
R&S
®
FSWP-B1
47
User Manual 1177.5862.02 ─ 09
Trigger 1/2
The trigger input and output functionality depends on how the variable "Trigger Input/
Output" connectors are used.
Note:
Providing trigger signals as output is described in detail in the R&S
FSWP User
Manual.
"Trigger 1"
"Trigger 1": "Trigger Input/Output" connector on the front panel
"Trigger 2"
Defines the usage of the variable "Trigger Input/Output" connector on
the rear panel.
"Input"
The signal at the connector is used as an external trigger source by
the R&S
FSWP. Trigger input parameters are available in the "Trig-
ger" dialog box.
"Output"
The R&S
FSWP sends a trigger signal to the output connector to be
used by connected devices.
Further trigger parameters are available for the connector.
Remote command:
OUTPut<up>:TRIGger<tp>:DIRection
Output Type ← Trigger 1/2
Type of signal to be sent to the output
"Device Trig-
gered"
(Default) Sends a trigger when the R&S
FSWP triggers.
"Trigger
Armed"
Sends a (high level) trigger when the R&S
FSWP is in "Ready for trig-
ger" state.
This state is indicated by a status bit in the
STATus:OPERation
reg-
ister (bit 5), as well as by a low-level signal at the "AUX" port (pin 9).
"User Defined" Sends a trigger when you select the "Send Trigger" button.
In this case, further parameters are available for the output signal.
Remote command:
Level ← Output Type ← Trigger 1/2
Defines whether a high (1) or low (0) constant signal is sent to the trigger output con-
nector (for "Output Type": "User Defined".