Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
92
User Manual 1177.5862.02 ─ 09
Example:
TRIG:SOUR EXT
Selects the external trigger input as source of the trigger signal
Manual operation:
See
See
See
See
See
See
9.4.2.2
Configuring the trigger output
The following commands are required to send the trigger signal to one of the variable
"TRIGGER INPUT/OUTPUT" connectors on the R&S
FSWP.
OUTPut<up>:TRIGger<tp>:DIRection
................................................................................ 92
...................................................................................... 92
....................................................................................93
OUTPut<up>:TRIGger<tp>:PULSe:IMMediate
.................................................................... 93
OUTPut<up>:TRIGger<tp>:PULSe:LENGth
........................................................................94
OUTPut<up>:TRIGger<tp>:DIRection
<Direction>
This command selects the trigger direction for trigger ports that serve as an input as
well as an output.
Suffix:
<up>
.
irrelevant
<tp>
Selects the used trigger port.
<2>: selects trigger port 2 (on the rear panel).
Parameters:
<Direction>
INPut | OUTPut
INPut
Port works as an input.
OUTPut
Port works as an output.
*RST:
INPut
Manual operation:
See
OUTPut<up>:TRIGger<tp>:LEVel
<Level>
This command defines the level of the (TTL compatible) signal generated at the trigger
output.
This command works only if you have selected a user-defined output with
.
Suffix:
<up>
.
1..n
Configuring MSRA measurements