Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
83
User Manual 1177.5862.02 ─ 09
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 15
Switches on 15 dB preamplification.
Manual operation:
See
INPut<ip>:GAIN[:VALue]
<Gain>
This command selects the "gain" if the preamplifier is activated (
INP:GAIN:STAT ON
,
The command requires the additional preamplifier hardware option.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<Gain>
For R&S
FSWP models 1322.8003K08, 1322.8003K09,
1322.8003K27 and 1322.8003K51, the following settings are
available:
15 dB and 30 dB
All other values are rounded to the nearest of these two.
For R&S
FSWP models 1322.8003K26 and 1322.8003K50:
30 dB
Default unit: DB
Example:
INP:GAIN:STAT ON
INP:GAIN:VAL 30
Switches on 30 dB preamplification.
Manual operation:
See
Scaling the Y-axis
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]
.......................................... 83
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
................................84
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:PDIVision
........................... 84
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:RPOSition
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
..........................................85
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]
<Range>
This command defines the display range of the y-axis (for all traces).
Configuring MSRA measurements