Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
88
User Manual 1177.5862.02 ─ 09
9.4.2
Triggering
The following remote commands are required to configure a triggered measurement in
a remote environment. These commands are only available for the MSRA primary
channel. More details are described for manual operation in
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
Configuring the triggering conditions
.......................................................................88
Configuring the trigger output
..................................................................................92
9.4.2.1
Configuring the triggering conditions
The following commands are required to configure a triggered measurement.
.............................................................................................88
TRIGger[:SEQuence]:HOLDoff[:TIME]
................................................................................88
TRIGger[:SEQuence]:IFPower:HOLDoff
............................................................................. 89
TRIGger[:SEQuence]:IFPower:HYSTeresis
.........................................................................89
TRIGger[:SEQuence]:LEVel[:EXTernal<port>]
.....................................................................89
TRIGger[:SEQuence]:LEVel:IFPower
................................................................................. 90
TRIGger[:SEQuence]:LEVel:IQPower
.................................................................................90
TRIGger[:SEQuence]:LEVel:RFPower
................................................................................90
............................................................................................ 91
..........................................................................................91
TRIGger[:SEQuence]:DTIMe
<DropoutTime>
Defines the time the input signal must stay below the trigger level before a trigger is
detected again.
Parameters:
<DropoutTime>
Dropout time of the trigger.
Range:
0 s to 10.0 s
*RST:
0 s
Default unit: S
Manual operation:
See
TRIGger[:SEQuence]:HOLDoff[:TIME]
<Offset>
Defines the time offset between the trigger event and the start of the measurement.
Parameters:
<Offset>
*RST:
0 s
Default unit: S
Configuring MSRA measurements