Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
100
User Manual 1177.5862.02 ─ 09
nel. The tasks for manual operation are described in
Chapter 6.9, "automatic settings"
Settings related to data acquisition (measurement time, hysteresis) can only be adjus-
ted in the MSRA primary application, not in the secondary applications.
[SENSe:]ADJust:CONFigure:LEVel:DURation
...................................................................100
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
........................................................ 101
..........................................................................................101
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
.............................................................. 101
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
............................................................... 102
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
●
Reference level
Example:
ADJ:ALL
Manual operation:
See
"Adjusting all Determinable Settings Automatically (Auto
[SENSe:]ADJust:CONFigure:LEVel:DURation
<Duration>
To determine the ideal reference level, the R&S
FSWP performs a measurement on
the current input data. This command defines the length of the measurement if
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
is set to
MANual
.
Parameters:
<Duration>
Numeric value in seconds
Range:
0.001 to 16000.0
*RST:
0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation:
See
"Changing the Automatic Measurement Time (Meastime
Configuring MSRA measurements