MSRA configuration
R&S
®
FSWP-B1
58
User Manual 1177.5862.02 ─ 09
Remote command:
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
Changing the Automatic Measurement Time (Meastime Manual)
This function allows you to change the measurement duration for automatic setting
adjustments. Enter the value in seconds.
Note:
The maximum measurement duration depends on the currently selected mea-
surement and the installed (optional) hardware. Thus, the measurement duration
actually used to determine the automatic settings can be shorter than the value you
define here.
Remote command:
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
[SENSe:]ADJust:CONFigure:LEVel:DURation
Upper Level Hysteresis
When the reference level is adjusted automatically using the
internal attenuators and the preamplifier are also adjusted. To avoid frequent adapta-
tion due to small changes in the input signal, you can define a hysteresis. This setting
defines an upper threshold that the signal must exceed (compared to the last measure-
ment) before the reference level is adapted automatically.
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
Lower Level Hysteresis
When the reference level is adjusted automatically using the
internal attenuators and the preamplifier are also adjusted. To avoid frequent adapta-
tion due to small changes in the input signal, you can define a hysteresis. This setting
defines a lower threshold that the signal must fall below (compared to the last mea-
surement) before the reference level is adapted automatically.
Remote command:
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
automatic settings