Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
82
User Manual 1177.5862.02 ─ 09
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
INP:ATT:AUTO ON
Couples the attenuation to the reference level.
Manual operation:
See
INPut<ip>:EATT:AUTO
<State>
This command turns automatic selection of the electronic attenuation on and off.
If on, electronic attenuation reduces the mechanical attenuation whenever possible.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
1
Example:
INP:EATT:AUTO OFF
Configuring a preamplifier
INPut<ip>:GAIN:STATe
<State>
This command turns the internal preamplifier on and off. It requires the optional pream-
plifier hardware.
The preamplification value is defined using the
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
Configuring MSRA measurements