Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
104
User Manual 1177.5862.02 ─ 09
Now you can send the
ABORt
command on the remote channel performing the mea-
surement.
Example:
ABOR;:INIT:IMM
Aborts the current measurement and immediately starts a new
one.
Example:
ABOR;*WAI
INIT:IMM
Aborts the current measurement and starts a new one once
abortion has been completed.
Usage:
Event
INITiate<n>:CONMeas
This command restarts a (single) measurement that has been stopped (using
ABORt
)
or finished in single measurement mode.
The measurement is restarted at the beginning, not where the previous measurement
was stopped.
, this command does not reset traces in
maxhold, minhold or average mode. Therefore it can be used to continue measure-
ments using maxhold or averaging functions.
Suffix:
<n>
.
irrelevant
Manual operation:
See
INITiate<n>:CONTinuous
<State>
This command controls the measurement mode for an individual channel.
Note that in single measurement mode, you can synchronize to the end of the mea-
surement with *OPC, *OPC? or *WAI. In continuous measurement mode, synchroniza-
tion to the end of the measurement is not possible. Thus, it is not recommended that
you use continuous measurement mode in remote control, as results like trace data or
markers are only valid after a single measurement end synchronization.
For details on synchronization see
.
If the measurement mode is changed for a channel while the Sequencer is active the
mode is only considered the next time the measurement in that channel is activated by
the Sequencer.
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
ON | 1
Continuous measurement
Capturing data and performing sweeps