Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
105
User Manual 1177.5862.02 ─ 09
OFF | 0
Single measurement
*RST:
1
Example:
INIT:CONT OFF
Switches the measurement mode to single measurement.
INIT:CONT ON
Switches the measurement mode to continuous measurement.
Manual operation:
See
INITiate<n>[:IMMediate]
This command starts a (single) new measurement.
You can synchronize to the end of the measurement with *OPC, *OPC? or *WAI.
For details on synchronization see
.
Suffix:
<n>
.
irrelevant
Manual operation:
See
INITiate:SEQuencer:ABORt
This command stops the currently active sequence of measurements.
You can start a new sequence any time using
Usage:
Event
Manual operation:
See
INITiate:SEQuencer:IMMediate
This command starts a new sequence of measurements by the Sequencer.
Its effect is similar to the
measurement.
Before this command can be executed, the Sequencer must be activated (see
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement is per-
formed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Manual operation:
See
Capturing data and performing sweeps