Applications and operating modes
R&S
®
FSWP-B1
24
User Manual 1177.5862.02 ─ 09
Sequencer State
Activates or deactivates the Sequencer. If activated, sequential operation according to
the selected Sequencer mode is started immediately.
Remote command:
Sequencer Mode
Defines how often which measurements are performed. The currently selected mode
softkey is highlighted blue. During an active Sequencer process, the selected mode
softkey is highlighted orange.
"Single Sequence"
Each measurement is performed once, until all measurements in all
active channels have been performed.
"Continuous Sequence"
The measurements in each active channel are performed one after
the other, repeatedly, in the same order, until sequential operation is
stopped.
This is the default Sequencer mode.
"Channel Defined Sequence"
First, a single sequence is performed. Then, only channels in continu-
ous sweep mode are repeated.
Remote command:
Refresh All
This function is only available if the Sequencer is deactivated, no sweep is currently
running, and only in MSRA mode.
The data in the capture buffer is re-evaluated by all active secondary applications, for
example after a new sweep was performed while the Sequencer was off.
Note:
To update only the displays in the currently active secondary application, use the
"Refresh" function in the "Sweep" menu for that secondary application (see
Remote command:
INITiate:SEQuencer:REFResh[:ALL]
Using the sequencer in MSRA mode