Applications and operating modes
R&S
®
FSWP-B1
23
User Manual 1177.5862.02 ─ 09
Remote command:
Replace Current Channel ← Selecting an MSRA secondary application
The secondary application selected on this tab is started in the currently displayed
channel, replacing the current secondary application.
Remote command:
Closing an application
To close an application, simply close the corresponding tab by selecting the "x" next to
the channel name.
Remote command:
4.3
Using the sequencer in MSRA mode
When you switch to MSRA mode, the Sequencer is automatically activated in continu-
ous mode. Unless it is stopped or you select a different Sequencer mode, the
R&S
FSWP continuously performs the following tasks:
1. The MSRA primary application acquires data.
2. The active secondary applications evaluate the data one after the other.
3. The display in the tab is updated.
This behavior is identical to "Signal and Spectrum Analyzer" mode (also for single
sequence or channel-defined Sequencer modes).
However, if you switch the Sequencer off, the behavior of the sweep functions is
slightly different to "Signal and Spectrum Analyzer" mode (see also
●
If continuous sweep is active (default) and you switch to a different secondary
application, continuous sweep is aborted. Thus, the same data is evaluated in dif-
ferent secondary applications without overwriting the data in the capture buffer. You
can start continuous sweep again as usual.
●
Only the secondary application that is currently displayed when a measurement is
performed is updated automatically. A new "Refresh" function is available to update
the display in one or all other secondary applications.
For details on the Sequencer function, see the R&S
FSWP user manual.
The "Sequencer" menu is available from the toolbar.
Using the sequencer in MSRA mode