Remote commands to perform measurements in MSRA mode
R&S
®
FSWP-B1
91
User Manual 1177.5862.02 ─ 09
Example:
TRIG:LEV:RFP -30dBm
TRIGger[:SEQuence]:SLOPe
<Type>
This command selects the trigger slope.
Parameters:
<Type>
POSitive | NEGative
POSitive
Triggers when the signal rises to the trigger level (rising edge).
NEGative
Triggers when the signal drops to the trigger level (falling edge).
*RST:
POSitive
Example:
TRIG:SLOP NEG
Manual operation:
See
TRIGger[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
Note on external triggers:
If a measurement is configured to wait for an external trigger signal in a remote control
program, remote control is blocked until the trigger is received and the program can
continue. Make sure that this situation is avoided in your remote control programs.
Parameters:
<Source>
IMMediate
Free Run
EXT | EXT2
Trigger signal from one of the "Trigger Input/Output" connectors.
Note: Connector must be configured for "Input".
RFPower
First intermediate frequency
(Frequency and time domain measurements only.)
IFPower
Second intermediate frequency
IQPower
Magnitude of sampled I/Q data
For applications that process I/Q data, such as the I/Q Analyzer
or optional applications.
BBPower
Baseband power (for digital input via the optional "Digital Base-
band" interface
PSEN
External power sensor
*RST:
IMMediate
Configuring MSRA measurements