FASTRAK MANUAL
Rev. G
66
JUNE 2012
‘R’ – *Reset Alignment Reference Frame
Syntax:
R
station
<>
Purpose:
This command resets the alignment reference frame for the specified station to the
station reference frame. It provides an easy way to re-align the reference frame to
the factory default values. The command parameter is defined as:
station
the number of the station to be reset.
Relatives:
A
Example:
Any time the alignment command (A1...) is used, it is best to send the reset
alignment command (R1<>) first. That way, there is no risk of building one
alignment on top of another. See
‘A’ – Alignment Reference Frame
for the Standard Alignment Procedure.