FASTRAK MANUAL
Rev. G
61
JUNE 2012
System Data Record: IEEE Floating-Point Format
RECORD IDENTIFIER ....... none
INITIATING COMMANDS ..... P or in continuous mode
byte(s)
Identification
Format
1 .....
Record type, “0”
A1
2 .....
Station Number
A1
3 .....
System error code
A1
(See Note 2 on previous page.)
?* .....
ASCII space character
A1
?* .....
Carriage return, line feed
?* .....
x,y,z position Cartesian coordinates
3(SingleFP)
?* .....
az,el,roll Euler orientation angles
3(SingleFP)
?* .....
X-direction cosines of the receiver’s x,y,z axes.
3(SingleFP)
(See Note 1.)
?* .....
Y-direction cosines of the receiver’s x,y,z axes.
3(SingleFP)
(See Note 1.)
?* .....
Z-direction cosines of the receiver’s x,y,z axes.
3(SingleFP)
(See Note 1.)
?*
The system data record contents are specified by the user using the
) and may vary from configuration to configuration.
Therefore, the specific location of a data item in the output record is not
determined until the record contents are defined.