‘c’ – Disable Continuous Printing
Syntax:
c
Purpose:
This command disables the continuous print output mode. After sending this
command, the continuous data stream from the FASTRAK to the host computer
will stop.
Relatives:
C, P
Default:
Continuous output mode is disabled
Example:
If the system is set to continuous output mode with the ‘C’ command, the user
may wish to stop the data stream to adjust other system parameters. This can be
accomplished by sending the command:
c