32 SCANNER EMC INFORMATION
168 Planmeca Chairside CAD/CAM solution
User’s manual
Rated maximum
output power
(W) of
transmitter
Separation distance (m) according to frequency of transmitter
150 kHz
to 80 MHz
80 MHz
to 800 MHz
800 MHz
to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.23
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended
separation distance (d) in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.