29 CLEANING / PREVENTIVE MAINTENANCE
User’s manual
Planmeca Chairside CAD/CAM solution 147
29.1.3 Autoclaving
Before autoclaving clean the tip as instructed in section
29.1.1 "Cleaning" above.
NOTE
Look for the temperature symbol engraved on
autoclavable tips. Only the tips with the temperature
symbol can be autoclaved.
1. Clean the surface of the mirror by using a cleaning tissue
or an optical cloth with a small amount of alcohol to gently
remove any residue. Dry the mirror with a dry cleaning
tissue, a gauze pad, or a dry optical cloth.
2. Individually pouch each scanner tip using an autoclave
pouch.
3. Place one to three pouches per tray.
4. Select the wrapped instruments cycle on the autoclave.
All tips can be autoclaved in 134°C autoclave cycle
according to the standard EN13060.
NOTE
Autoclave cycle names differ among manufacturers.
Reference your manufacturer’s manual to select the cycle
for sterilizing wrapped instruments.
5. Autoclave the pouched tip.
6. Store in pouch for later use.
7. Prior to scanning check that the scanning tip mirror is
clean and dry. If necessary you can clean the surface of
the mirror as instructed in step 1 above.
29.2 Cleaning the system
NOTE
For intraoral scanning systems only.
NOTE
Protect the keyboard with a disposable barrier.
NOTE
Before and after each use, clean all areas of the scanner.
WARNING
Before and after each use, follow these instructions to
disinfect the PlanScan scanner. Do not substitute any
other cleaning solution or procedure. Never use any
paint thinner, solvents, or harsh chemicals.
NOTE
Use only a non-woven sponge or pre-moistened
germicidal cloths that have been saturated with a hospital
grade disinfectant when cleaning the PlanScan scanner.