If you want to calibrate the HID's axes directly afterwards, proceed as
.
■
If you want to set up and activate HID control for the E-871.1A1N directly afterwards,
close the
Test and Calibrate HIDevice
window with
Close
and continue as described in
"
Setting up and Activating HID Control (p. 91)
".
■
If you do not want to make any further settings at this point, close
Test and Calibrate
HIDevice
with
Close
and the
Configure Human Interface Devices
window with
OK
.
Calibrating HID Axes
1. If necessary, open the window for configuring HID control via the
E-871.1A1N > Configure
controller HIDevice(s)…
menu item in PIMikroMove's main window.
➔
The
Configure Human Interface Devices
window opens.
2. If necessary, open the window for calibrating the HID and click the
Test and calibrate...
button.
➔
The
Test and Calibrate HIDevice
window opens.
3. Select the
Calibrate axes
tab in the
Test and Calibrate HIDevice
window.
4. Select the respective lookup tables to be used for the HID's axes via the selection fields in
the
Calibrate axes
tab.
➔
The example shows that a user-defined lookup table was selected for axis 1.
8 E-871.1A1N FUNCTIONS
PZ285EN
‒
1/30/2019
89