6
Installation
6.1
Mounting the E-871.1A1N
The E-871.1A1N can be used as a benchtop device or mounted on a surface in any
orientation or installed in a control cabinet.
Overview
5
5
4
43,9
21,95
4,5
12
1
127
M4
147
5
5
60
22,5
104,5
4,5
9,8
2,2
Area
Revision
Revision Note
ALLGEMEINTOLERANZ
STANDARD TOLERANCE
BEHANDLUNG / TREAT
RAUHEIT / FINISH
DATE
NAME
BEZEICHNUNG / TITLE
All information provided in this drawing is to be
considered confidential, proprietary and secret
information. It is intendedonly for the use of the
individual or entity to whom it is sent. Individual
or entity agrees to not disseminate or copy this
information to anyone other than originally
specified without the written consent of
PHYSIK INSTRUMENTE GmbH & Co.KG
Tolerierung ISO 8015
DIN ISO 2768
WERKSTÜCKKANTEN / WORKPIECE EDGES
m - K
Ra
Figure 3: Dimensions and position of the recesses for mounting
Tools and Accessories
■
Suitable screws
■
Suitable screwdriver
Requirements
✓
You have read and understood the
general safety instructions (p. 11)
.
NOTICE
Heating up of the E-871.1A1N during operation!
High temperatures can overheat the E-871.1A1N.
►
Install the E-871.1A1N with a gap of at least 10 cm to the top and rear panels and at least
5 cm to its sides. If this is not possible, make sure that the surroundings are cooled
sufficiently.
►
Ensure sufficient ventilation at the place of installation.
►
Keep the ambient temperature at a noncritical level.
Mounting the E-871.1A1N on a Surface
1. Drill the holes required into the surface.
6 INSTALLATION
PZ285EN
‒
1/30/2019
44