Configuring the System
8-18
Start-up and Configuration
Performance:
Perform an operational test of the Clinical Network by
executing a data passing operation from each connected
Information Center to every other Information Center,
Client, or Printer on the Network
For each M3150 and M3155 Information Center:
Verify that a review application (e.g.
Wave Review
) can
be executed. This verifies connection to the M3154
Database Server (if a M3155 Information Center).
For each M3151 Client:
Verify that waveforms from every Information Center on
the Network having a connection to an SDN can be
viewed on the Client.
Verify that a review application (e.g.
Wave Review
) can
be executed. This verifies connection to the M3154
Database Server
For each wireless Patient Monitor:
Verify that waveforms from each wireless patient monitor
on the network are displayed on an Information Center
and that the waveforms are continuous.
Expected answers are
“Yes”. If so,
Performance
test passed.
P:P
or
P:F
where
P
= Pass
F
= Fail
Safety
No safety test is required
S:NA
where
NA
= Not
required
Table 8-7. M3185 Clinical Network Components - Test and Inspection Matrix
Test Block
Name
Test or Inspection to Perform
Expected Results
What to Record
on Service
Record
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...