Configuration
Software Installation & Configuration
6-21
Warning
If
Hemodynamic Pressures
and/or
Gas Pressures
are changed, the entire patient
database for the Server -- Alarms, Waves, Trends, and Parameters -- will be erased.
Local General Config Settings
set at the individual Information Center or Client are only
applicable to that particular device.
Warning
The height, weight, and gas units of measure configured on the Information Center must match the units of
measure at the bedside. Failure to do so may result in incorrect hemodynamic calculations and patient
documentation.
Multi-lead ECG Layout
Sets several initial presentation parameters for the display in the 12-lead review applications. The
first set of selections sets the type of layout of ST segment review data -
International
or
Cabrera
. The
Layout
group selects the layout for the presentation of the 12-lead waveforms, as
either
3 rows by 4 columns, 6 rows by 2 columns
, or
12 rows by 1 column
. The
Speed
group allows for selecting the chart speed used in presenting the waveforms, and can be either
25
mm/sec
or
50 mm/sec
.
Patient Name in
Overview
If
yes
, the patient name for all overviewed beds is displayed in the alarm reflector, overview bed
list, overview window, and alarm overview window. If
no
, the name is not displayed. This setting
is for M2/M3/M4/IntelliVue patient monitors monitored by the Information Centers connected to
this Database Server. This setting does not apply to SDN beds.
Allow patient
navigation for Web
Enabled only if the Patient Data/Transfer parameter was enabled in the Purchased Options
window.
Note—
This option must be disabled if automated web access is used.
Web Overview Port
Port of the TCP/IP socket connection for web access. The default is
9000
.
Field
Description
Field
Description
Patient Type
Default setting for the type of patient being monitored --
Adult
,
Pediatric
,
Neonate --
for
SDN
Bedside
monitors. This can be changed for specific patients in the
Admit
window.
All Arrhythmia Alarms
Off (for SDN Bedsides
Only)
Arrhythmia alarms from telemetry monitors can be turned off in the
Arrhythmia Alarms Setup
menu for
SDN Bedside
monitors.
Enabled
means they can be turned off.
Disabled
means
they cannot be turned off.
No Data From Bed
INOP - alerting
INOP
sound is to be sounded when there is a NoDataFromBed INOP. The
NoSound
selection
disables the sound alert while the
Sound
selection will play the alert tone
Arrhythmia Capability
Sets whether the
Enhanced
or
Basic
set of arrhythmia measurements are to be used for
SDN
Bedside
monitors.
Enhanced
has 22 arrhythmia alarms.
Basic
has 10 arrhythmia alarms
Note
The Arrhythmia Capability parameter configuration (
Enhanced
or
Basic
) must have the same configuration
setting enabled in the Patient Monitor (if arrhythmia is on).
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...