Network Names and IP Addresses
5-44
Hardware Installation
Step 2.
Select
Accessories
to show the
Accessories
menu.
Step 3.
Click on
Command Prompt
to bring up the
MS DOS Console
window shown
Figure 5-39 MS DOS Console Window for Testing Network Connectivity
Step 4.
Type
ping hostname
(or
ping IPAddress
) for one of the Network connected
Information Centers after the
C:\ >
prompt, as shown for Host Name =
HPICU
(and
HPCCU
, and press
Enter
.
If the ping is successful, i.e. the network connection is complete and the Host Name is
correct, a reply using the device’s correct IP Address will appear, as shown in
.
If the ping fails, the reason for the failure (e.g. bad IP Address) will be shown. Identify the
problem, correct it (e.g. use an IP Address instead of the Host Name), and repeat the process.
Step 5.
Repeat
Step 4
for each Information Center, Client, Switch, Access Point, and
Printer on the Network.
Step 6.
Click
Yes
to restart the computer.
Note
Allow the computer to fully reboot before proceeding.
Note
Allow the computer to fully reboot before proceeding.
After the
Host Name
has been changed,
Philips software must be reinitialized
. See
Installation Procedure
.
Step 7.
Repeat this process for each Information Center, Client, and the Server.
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...