Troubleshooting
7-30
Maintenance, Troubleshooting, and Repair
dataserver
redirector
recmgr
sdprocess
pdsstore
docstatusserver
Error: Unhandled
Exception Filter is
kicking in
Fatal error. System will automatically
reboot to clear error state
Reboot system to clear. Contact Response
Center if it happens often.
EMSMainMenu Winsock connection failed
Reboot system. If this persists, reinstall OS
software.
Mode of operation could not be read from registry. Therefore,
the mode was defaulted to monitoring mode.
No action required.
The user logon flag could not be read from the registry.
Therefore, the user logon was defaulted to the
SystemManager account
Reboot system
Could not open the specified event to send to the Shell
application.
Reboot system
Could not start application because Error = XX
Reboot system. If this persists, reinstall
Philips software
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Reboot system. If this persists, reinstall OS
and Philips software
Could not write value to Registry because Error = XX
Reboot system. If this persists, reinstall
Philips software
Install
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Reboot system.
Could not write value to Registry because Error = %d
Reboot system. If this persists, contact
Response Center
OpenProcessToken failed.
Contact Response Center
AdjustTokenPrivileges enable failed.
Contact Response Center
Auto Reboot Failed
Reboot system or recycle power
NetCfgLib
Auto Reboot failed
Reboot system, If this persists, reinstall
Philips software
OpenProcessToken failed
Reboot system. If this persists, reinstall
Philips software
AdjustTokenPriveleges enable failed
Reboot system. If this persists, reinstall
Philips software
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Reboot system. If this persists, reinstall OS
and Philips software
Could not write value to Registry because Error = XX
Reboot system. If this persists, reinstall
Philips software
Could not read value from Registry because Error = %d
Reboot system. If this persists, reinstall
Philips software
Could not write to Error File
No action required
OrbPlus
Error: all
No action required
Table 7-7. Application Events
Source
Description
(Event Code)
Explanation
Corrective Action
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...