Configuration
6-22
Software Installation & Configuration
Screen Notes
Screen Notes
made in the patient
Admit
window will be displayed on the
Main Screen
.
Yes
means they will be displayed.
No
means they will not be displayed.
Alarm Sounds
• CareNet
• IEC
Note—
IEC alarm sounds are only supported on IntelliVue and TeleMon Patient Monitors. V24
and CMS monitors only support Traditional (CareNet) alarm sounds. If the Information Center
has a mix of these monitor types, configure the Information Center to use CareNet in order to
have consistent alarm sounds across these monitoring devices. Verify all the monitors are
configured for the Traditional (CareNet) alarm sounds also.
Record Button
Buttons for recording and/or storing waveforms display on the
Main Screen
•
Record
means that a button for recording waveforms will be shown.
•
Store
means that a button for storing waveforms will be shown.
•
Both
means that buttons for both recording and storing waveforms will be shown.
Silence Bed at Central
Button for silencing bedside monitor alarms at the device appears in
Patient Sectors
.
Yes
means that a silencing button appears.
No
means it does not appear and alarms can only be
silenced at the bedside monitor.
Web Access Portal
If enabled, the
Browser
button is present on the All Controls window. The
Browser
button is
used to open a web browser for viewing browser based applications when an Application Server
is connected.
HL7 Export
Configuration
Active only if HL7 Export option is selected in the Purchased Options and Support Information
page.
When selected, a HL7 Configuration window opens:
• LAN Based Monitors only:
Enable
or
Disable
.
If enabled, only data originating from
LAN monitors (M2/M3/M4 or IntelliVue Patient Monitors) is sent.
• Send Alert Messages:
Enable
or
Disable
.
If enabled, alert, INOP, and parameter data is
output in HL7.
• Send Aperiodic Data:
Enable
or
Disable
.
If enabled, aperiodic measurements (e.g.
noninvasive blood pressure) is included in the HL7 output.
• Send Patient Name:
Enable
or
Disable
.
If enabled, the patient name is included in the
HL7 output. This may not be allowed per hospital policy.
• Target: Specifies the HL7-client host name that the HL7 data is to be exported to.
Field
Description
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...