Configuring the System
8-8
Start-up and Configuration
•
Use the ruler to measure (in millimeters) the full width (
W
) of a waveform in a
Patient
Sector
of the
Main Screen
display as shown below. Use the ruler measurement
techniques of
Step 5 of
•
Measure (in seconds) the time (
t
) for 10 full sweeps of the erase bar for the same
waveform as shown below. The measurement shall be to the nearest second.
•
To meet the sweep speed specification of 25 mm/s +10%, the measured time (
t
) shall be
between:
Modem Test
The following procedure tests a factory installed modem on a PC or server and external dial-
in access.
Step 1.
Click
Start
in the
Windows Main Menu
and then
Programs
.
Step 2.
Click
Accessories
in the
Programs
menu.
Step 3.
Click
Hyperterminal
in the
Accessories
menu.
Step 4.
Click
HyperTerminal
in the
Hyperterminal
menu.
If the
Location Information
window of
appears:
•
Select a
country/region
•
Enter an
area or city code
•
Select whether the phone is
Tone dialing
or
Pulse dialing
0
Width
(
W
)
Waveform Width in Millimeters (
W
)
Total Time of 10 Sweeps in Seconds (
t
)
(
t
)
10
9.1 x
W
11.1 x
W
25
25
<
t
<
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...