Configuration
6-56
Software Installation & Configuration
Time & Date
Setup
Time & Date Setup
is used to set and synchronize the date and time on all Information
Centers and Clients on the Network and all patient monitors connected to SCCs. It can be set
in monitoring mode on the Server or any Network connected Information Center or Client.
Settings are automatically stored on the Database Server and distributed to all other
Information Center and Clients on the Network within seconds. The
Time and Date Setup
window is shown in
Caution
In order for the time and date to be set on SDN patient monitors, one Information
Center must be connected to Branch 0 of each SCC.
Figure 6-17 Time and Date Setup Window
Description
Fields
in the
Time and Date Setup
window are:
Notes
In some languages, Time Zone will not automatically adjust for daylight savings correctly. If
this happens, Time Zone must be reset manually.
Field
Description
Date: (MDY)
Sets the current
M
onth
,
D
ate
, and
Y
ear.
To change the values, do the following:
• Click in a box to highlight the number.
Use the up/down arrows to the right of the field to increase or decrease the value in the box
Time:(H:mm)
Sets the current
H
our
and
minute (
mm
)
. To change the values, use the same procedure as above
Time Zone Information
Is for setting the following
•
Time Zone
sets the correct
Time Zone
for the system location
The procedure is as follows:
• Click the
arrow
to the right of the
Time Zone
field to display the list of
Time Zones
.
• Click on (highlight) the correct
Time Zone
for your location.
Automatically Adjust
For Daylight Savings
Time
Selects whether the time will automatically adjust when the time changes to or from
Daylight
Savings Time
.
• Clicking a check in the box
causes time to automatically change.
• No check
means it will not change.
Store Settings
Activated
only
if
Date and Time Setup
is opened directly from the
Configuration
menu of
the
Service
application and
not
from
Config Wizard
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...