Troubleshooting
Maintenance, Troubleshooting, and Repair
7-31
ProductConfig
Could not open HKEY_LOCAL_MACHINE Registry
because Error = XX
Reboot system. If this persists, reinstall OS
and Philips software
Could not write value to Registry because Error = XX
Reboot system. If this persists, reinstall
Philips software
Could not read value from Registry because Error = %d
Reboot system. If this persists, reinstall
Philips software
WinSock connection failed
Reboot system. If this persists, reinstall OS
software
Could not read which configuration applications to run from
the Registry.
Reinstall Philips software
Could not start application because Error = XX
Reboot system.
PswdScrDlg
Could not start EMS Main Menu application
Reboot system. If this persists, reinstall
Philips software
... Application could not be started.
Reboot system. If this persists, reinstall
Philips software
Logon User does not have the privilege to access EMS
applications. Need to update the access rights of the
SystemManager User.
Reboot system. If this persists, reinstall
Philips software
WinSock connection failed
Reboot system. If this persists, reinstall OS
software
System manager password entered
Verify proper user access
Philips Private password entered
Verify proper user access
RecMgr
Error: 2504, The default case on a switch was reached.
Report to Response Center and reboot
system.
Error: 2510, No primary wave was found in the pds. Using the
first two waves.
Reboot system. If this persists, reinstall
Philips software.
Error: 2511, No data was found in the pds. This could be a
temporary startup condition.
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
Error: 2512, A pds collector could not be created for the
specified patient.
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
Error: 2513, A patient could not be found from the
information supplied.
If this occurs around a patient
discharge or transfer or when a
bedside monitor module is pulled
out, no action required.
If occurs at other times, reboot system.
Error: 2514, This unmarshall routine does not support the
version of data that was marshalled.
Reboot system. If problem persists, reinstall
Philips software. If problem still exists,
contact Response Center.
Table 7-7. Application Events
Source
Description
(Event Code)
Explanation
Corrective Action
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...