System Components
Hardware Description
2-21
4-Channel recordings are routed to the 4-channel recorder. The recording destination is set
either by the Configuration Wizard control or by the Recording Setup application (under Unit
Settings), separately, for the 4-channel recorder and for the 2-channel recorder. Refer to the
Recorder Destination Screen in the Configuration Wizard section of this manual. The default
setting for this control is the local machine if a 4-channel recorder is installed. Otherwise, it is
set to none, which disables any control that requests a 4-channel recording. When a
Information Center is fully configured the User can select any Information Center with a 4-
channel recorder as a destination.
The M3160 4-Channel Recorder has the following selectable recorder
speeds
. Recorder speed
accuracy
is + 2%.
50 mm/s
300 cm/min.
25 mm/s
150 cm/min.
12.5 mm/s
75 cm/min.
The Recorder print head has the following characteristics:
Width:
104 mm printable width
Resolution:
8 dots/mm x 8 lines/mm
System Behaviors with the 4-Channel Recorder:
Error Messages
- the Information Center resting display status line will display messages
that indicate any error conditions with the recorder (e.g.,
4-Channel Recorder Not Ready
which indicates out-of-paper, or no power, or not connected, or internal failure).
Queuing
- if the user requests a 4-channel recording and the recorder is already busy, then the
recording will be held in queue until the recorder becomes available.
Gain Change
- whenever the gain is changed for a waveform in an on-going 4-channel
recording, the wave label is annotated on the recording strip.
Remote Access Service
- RAS is no longer supported when the 4-channel recorder is used on
Information Centers unless accessed via a database server MODEM.
BIOS Settings
- Changes may be required to the BIOS settings when a 4-channel recorder is
added. Refer to “Updating BIOS” on page 7-119.
Figure 2-22 Philips 4-Channel Recorder Rear Panel Connections
4 - Channel Recorder
24V Power Supply
to Serial B Port Interface
Cable to PC Workstation
to UPS
AC Output
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...