Troubleshooting
Maintenance, Troubleshooting, and Repair
7-89
ST/AR Configuration Reporting Tools
This section provides a list of the 7-character encoded ST/AR configuration parameters that
display as part of the alarm information on alarm strip recordings. This information can be
used to identify problems in the operation of ST/AR arrhythmia analysis as well as in the
interpretation of results obtained by the algorithm. The information is encoded with a [ ]
after the time stamp.
Note
The ST/AR diagnostic tool encoding scheme is only available on SDN and Telemetry beds.
The items are coded into a 7 character string, broken down as follows:
First Character
Codes
The first character identifies the ST/AR revision.
Character
Identifies
First
ST/AR’s revision.
Second
Patient category, pacing mode and analysis level.
Third and Fourth
• Classification mode (single or multi-lead)
• Detection mode (auto or manual)
• User specified lead label (manual detection mode only)
• User specified minimum threshold (150-350uV) (manual detection
mode only)
• Algorithm minimum detection threshold (150-350uV) (manual
detection mode only)
Fifth
Number of active classification and detection channels
EASI mode/EASI coefficient set
Sixth
ECG hardware source and lead set in use (3-wire, EASI, etc.)
Seventh
Source of Asystole, Pause, Missed Beat, PNP or PNC alarm (detection or
beat interval)
STAR revision
First Character
Release E
4
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...