Symptom
Cause
Remedy
Pressure
The unit will not
hold pressure
The cell o-rings are worn or
not seated properly
There is a leak at one of the
fittings
If there are no visible
leaks then there may be
an internal leak caused by
clogged tubing or a faulty air
pressure valve.
Disassemble the cell and
inspect all o-rings. Discard
any that show signs of
damage or wear
Tighten the leaking
connection
Inspect filters as instructed
on page 66 and replace
if needed. If the tubing is
clogged or the air pressure
valve has failed, contact
OFITE Technical Support.
The unit will not
build pressure
There is a leak at one of the
fittings
The pump is broken (not
cycling, constantly cycling)
The Rupture Disk is blown
If there are no visible
leaks then there may be
an internal leak caused by
clogged tubing or a faulty air
pressure valve.
Tighten the leaking
connection
Replace the pump
Replace the Rupture Disk
as instructed on page 66.
Inspect filters as instructed
on page 66 and replace
if needed. If the tubing is
clogged or the air pressure
valve has failed, contact
OFITE Technical Support.