4. Start the SGSM software.
5. Put the “UCA/SGSM” switch to the SGSM position.
6. Click Utilities
→
Calibrate Load Cell.
7. You will be prompted to take a reading with no weight on the transducer.
Simply click the “ACCEPT” button.
8. You will now be prompted to add weight to the weight holder. Place the
appropriate weight on the holder and click the “ACCEPT” button.
When calibrating the unit for the first time, the software does not provide
prompts that indicate which weights to apply to the weight holder.
The correct weights are: 0g, 100g, 200g, and 400g.
9. Continue adding weight according to the prompts and click “ACCEPT”
every time.
10. When the calibration is complete, click the “SAVE” button.
11. Remove the calibration pulley assembly and reinstall the tension spring
between the base spring post and the torque spring post.