h. The next screen will display the new cell values. Click “Save” to store
the new value.
i. Pull the calibration shunt jumper. Notice the red flaps are visible
again.
Red Flaps
Visible
The Pressure Transducer is now calibrated.
f. Wait for the value to stabilize.
g. Once the value stabilizes, click “Accept.”