44
SGSM
Cell Cap Assembly
The cell body and both cell caps were manufactured and pressure tested
together. All three pieces are serialized. Before assembling the test cell, make
sure all three pieces have the same serial number.
For a complete diagram of the test cell, refer to page 85.
Bottom Cap Assembly
1. Inspect the transducer port to make sure that it is clean and free of debris.
Dried transducer couplant can accumulate on the sides and bottom of the
port. This couplant must be cleaned from the port as well.
2.
Tighten the jaws of the vice on the flats of the cell cap with the threads
facing up.
3. Apply a thin layer of high-temperature grease to the surface of the bottom
cell cap that meets with the cement. Wipe off any excess grease.
Bottom Cell Cap
High Temp Grease
Application Area