d.
Hold Indefinitely (For UCA only):
With this option turned on, the test will run until you click the “Stop
Test” button. If this option is not checked, the test will end when all the
steps in the Temperature Profile are complete.
In the SGSM Profile Builder, there is no “Hold Indefinitely” option. The
test will continue to hold temperature through the final gel strength
until the end of the test.
5. Click the “OK” button to exit the “Setup” screen. The new Temperature
Profile will now appear in the “Temperature Profiles List” in the “UCA Info”
tab.
b. Pressure – There are two types of pressure settings:
i. Hold – This will hold the current pressure for a set number of
minutes.
ii. Ramp – This will increase the temperature up to the target in a set
number of minutes. Enter the time to temperature in the "Time"
box. Enter the target temperature in the "Target" box.
iii. Step – This will increase the pressure up to the target as fast as
possible. Enter the target temperature in the "Target" box.
The maximum pressure setpoint allowed is 20,000 PSI (138 MPa).