OFITE, 11302 Steeplecrest Dr., Houston, TX 77065 USA / Tel: 832-320-7300 / Fax: 713-880-9886 / www.ofite.com
67
At times moisture can build up through out the air lines if the air supply into
the machine is not dry. The air filter will indicate if moisture is building up
through out the lines. If so, the water should be removed from the air line and
the air filter should be drained of excess moisture.
Pressure Release Button
1.
Place a paper towel underneath the air filter to prevent water from
spraying inside the UCA/SGSM.
2. Use the paper towel to press the pressure release button to remove water
in the line.