Stopping the Test
1. The test stops automatically. The alarm needs to be acknowledged and
cleared by clicking the SGSM alarm (green button) that appears in the
upper right hand corner.
2. Turn the “HEATER” switch the "OFF" position. If the cooling system is
switched to the “Auto” position, the cooling system will cool automatically
if the cell temperature is greater than 100°F. To change the temperature
threshold, contact OFITE Technical support.
3. Turn the “COOL” switch to the “MANUAL” position and allow the test cell
to cool to 100°F.
4. When the cell has cooled, turn the “PRESSURE UP” and “COOL”
switches to the "OFF" position. Turn the “WATER ON/FILL” switch to the
“OFF” position.
Always open the pressure release valve very slowly to prevent
pulling cement into the plumbing.
5. Unscrew and remove the high-pressure line from the cell cap and the
instrument cabinet.
Use caution. The test cell may contain a small amount of air at high
temperature and/or pressure.
6. Unplug the thermocouple.
7. Remove the SGSM cap assembly.
8. Remove the thermocouple.
9. Lift the cell out of the heating jacket.